Κοινά σφάλματα και λύσεις του αερίου ανιχνευτές

 

Σφάλμα 1: Ανίκανος να ανίχνευση at low συγκέντρωση


1. If it is a pump suction measurement, first check if the pump is working. when the pump is working normally, there is slight vibration, and you can feel important suction by blocking the air inlet with your fingers for 5 seconds. Then check if the air inlet is blocked or the connection is not correct sealed, resulting in air leakage and no suction.


2. Έγχυση άζωτο αέριο σε βαθμονόμηση το μηδέν σημείο ή βαθμονόμηση το μηδέν σημείο in καθαρός αέρας, και αμέσως εκτέλεση δοκιμή μετά βαθμονόμηση.


3. After calibrating the zero point, the measured gas cannot be detected, and it is necessary to restore the factory settings.


4. Μετά αποκατάσταση το εργοστάσιο ρυθμίσεις, it cannot be detected. It is necessary to inject nitrogen again or perform zero calibration in clean air. After calibration, the test should be carried out αμέσως.


5. Check if the connection wire of the sensor has been artificially damaged or has poor contact.


6. The above four steps have been completed but cannot be detected. it is necessary to confirm if there is any gas being measured on site, or if the concentration of the gas being measured is indeed very low. if it is below the minimum detection limit of the instrument, it cannot be detected.


Fault 2: In the air, there is no measured gas, but the value fluctuates greatly or jumps randomly

Λύση%3α


1. Γενικά, a short time zero point fluctuation range less than 1 percent of the maximum range is seen a normal range, and a long time drift less than the maximum range without the measured gas is seen a normal range

2 ποσοστό είναι εντός του κανονικού εύρους. Αν αυτό υπερβαίνει αυτό αυτό εύρος%2γ είναι απαραίτητο να επιβεβαιώνει αν υπάρχει υπάρχει οποιοδήποτε αέριο είναι μετράται επιτόπου, ή αν το θερμοκρασία και υγρασία σε ο αέρα διακύμανση σημαντικά, αποτέλεσμα σε ασταθής τιμές. Γενικά, θερμοκρασία και υγρασία διακυμάνσεις μπορεί αιτία βραχυπρόθεσμα και σημαντικά διακυμάνσεις στο όργανο's ανίχνευση τιμές. Μετά το θερμοκρασία και υγρασία στο ο αέρας παραμένει σταθερό% 2c the values will also be relatively stable.


2. Confirm whether the instrument has undergone zero calibration or target point calibration. If zero calibration is performed in the presence of the measured gas, low concentration gas may not be detected. If target point calibration is performed in the presence of the measured gas, but the calibrated concentration value does not match the actual concentration value, it may cause significant fluctuations in the instrument value or the detected value to be too small, Both cases can be resolved by restoring the factory operation.


3. If the problem cannot be solved after restoring the factory settings, it is necessary to confirm whether the power supply of the detector is sufficient. Do not use a transformer to power the instrument. Instead, use a DC stabilized power supply or switch power supply, and the power should be greater than 24V, 1A (for single device power supply). Multiple devices require higher power, and the wire diameter of the power cord should not be too small, The wire diameter for short distance power supply to a single device is better than 0.75 square millimeters.


4. Check the routing of the signal transmission line to see if it passes through high-power fans, motors, frequency converters, and other areas that may cause electromagnetic field changes. Try to avoid areas with strong electricity and electromagnetic fields as much as possible, and use shielded wires as the signal transmission line. One end of the shielding layer should be connected to the instrument casing, and the other end θα πρέπει να είναι γειωμένο να εξασφαλίζει καλό γείωση.

 

Fault 3: Inaccurate detection

Λύση%3α


1. Επιβεβαίωση εάν το αέριο συγκέντρωση στο τοποθεσία είναι ακριβές. Μερικές φορές, υπάρχει α σημαντική διαφορά μεταξύ το θεωρητικό τιμή και το πραγματικό τιμή. Αυτό είναι καλύτερο να επαληθεύει το ακρίβεια του του μέσο με εισάγοντας πρότυπο αέριο ή αποστολή αυτό σε α τρίτος μέτρηση ίδρυμα για δοκιμή.


2. If the sensor is used for a long time, there may be some errors in the measured values. It is necessary to confirm with the manufacturer if the sensor can still be used. If the sensor itself is approaching its its service life, even if it it be used normally after calibration, it is recommended to replace the sensor.

 